The MiniMed Flex Insulin Pump system has SmartGuard technology, which continuously helps protect against high or low glucose levels. Its Meal Detection technology helps cover for underestimated carb counts or an occasional missed meal bolus. The MiniMed Flex insulin pump pairs with the MiniMed Simplera Sync sensor and allows for glucose targets of 100, 110 and 120 mg/dL.
This product is for the management of diabetes in persons requiring insulin who are ages 7 years old and older.
For more information, visit MiniMed.
Activity Settings:
Temporary basal rate can be set, and multiple basal rate profiles can be set.
Automatic Correction Boluses:
Yes
Closed Loop Functionality:
Yes
Automatic Insulin Adjustments:
Yes
Calibration Not Required:
Yes
Compatible CGM:
Yes, compatible with Simplera Sync CGM sensor.
Waterproof:
The MiniMed Flex pump, while used with an infusion set and an insulin reservoir, is protected against water damage if the water is a maximum of 8 feet deep and the time under water does not exceed 105 minutes.
Medicare Covered:
NA yet
Remote Software Update Capability:
Yes
Multiple Languages:
No
Food Database:
No
Additional Features:
Size Weight:
Size is 1.0 inches W x 3.8 inches L x 1.4 deep. Weight is 73 grams.
Approved For Ages:
Indicated for persons ages 7 and above
Tubeless:
No
Battery:
Internal battery charged using a dedicated charging base with a USB-C cable and wall plug. Full charging takes about 30 minutes.
Reservoir Capacity:
300 units
Infusion Set:
Compatible with all Medtronic infusion sets including the Extended Infusion set which lasts up to 7 days
Basal Range:
Minimum basal insulin increment: 0.050 units/hour.
Above 0.050, increments can be set in 0.025-unit increments (e.g., 0.05, 0.075, 0.1, etc.).
Maximum hourly basal rate is 15 units.
Bolus Range:
Minimum bolus dose of insulin is 0.05 units. The default increment is 0.1 units but can be reduced to .01 units. Maximum bolus is 25 units.
Transmitter Range:
Maximum of ten feet between pump, sensor, and mobile device.
Danatech Resources
MinMed Resources
For complete details of the system, including product and important safety information such as indications, contraindications, warnings and precautions associated with the system and its components, please consult https://www.minimed.com/en-us/important-safety-information and the MiniMed Flex user guide.
DISCLAIMERS:
This site and its services do not constitute the practice of medical advice, diagnosis or treatment. Always talk to your diabetes care and education specialist or health care provider for diagnosis and treatment, including your specific medical needs. If you have or suspect that you have a medical problem or condition, please contact a qualified health care professional immediately. To find a diabetes care and education specialist near you, visit the ADCES finder tool.